Terminology compendium

Every manufacturer calls the same concept something different. The naming table maps each setting across the three devices, and the dictionary below defines the terms.

Same setting, three different names

Tidal volume
ReVelVt
ZOLL Z VentVt
Hamilton T1Vt/IBW

Hamilton sets volume per ideal body weight

Respiratory rate
ReVelRate
ZOLL Z VentBPM
Hamilton T1f

ZOLL counts breaths per minute; Hamilton uses 'f' (frequency, b/min)

PEEP
ReVelPEEP
ZOLL Z VentPEEP
Hamilton T1PEEP/CPAP

Combined label on Hamilton

Oxygen concentration
ReVelFiO2
ZOLL Z VentFIO2
Hamilton T1O2

ZOLL writes FIO2; Hamilton shows O2 as a percentage

Inspiratory time
ReVelI-time
ZOLL Z VentTi
Hamilton T1TI

The Z Vent calls it the control parameter — I:E or Ti

Peak inspiratory flow
ReVelFlow
ZOLL Z VentFlow
Hamilton T1Insp Flow
Pressure support
ReVelPS
ZOLL Z VentPS
Hamilton T1ΔPsupp
Inspiratory pressure
ReVelPC
ZOLL Z Vent—
Hamilton T1ΔPinsp

The Z Vent sets its pressure target as the Mode window's breath target; Hamilton's is above PEEP

Trigger / sensitivity
ReVelSensitivity
ZOLL Z VentTrigger
Hamilton T1Trigger

Flow trigger in L/min on the Z Vent and Hamilton

Minute volume
ReVelMV
ZOLL Z VentMV
Hamilton T1ExpMinVol

Expired minute volume

Exhaled tidal volume
ReVelVte
ZOLL Z VentVte
Hamilton T1VTE
Peak pressure
ReVelPIP
ZOLL Z VentPIP
Hamilton T1Ppeak
Plateau pressure
ReVelPplat
ZOLL Z VentPplat
Hamilton T1Pplateau
I:E ratio
ReVelI:E
ZOLL Z VentI:E
Hamilton T1I:E

Displayed on all three

Backup rate
ReVelBackup Rate
ZOLL Z VentBackup BPM
Hamilton T1f backup

NIV/timed modes

Volume mode name
ReVelA/C VC
ZOLL Z VentAC (v)
Hamilton T1(S)CMV+

Assist-control with a volume target

Pressure mode name
ReVelA/C PC
ZOLL Z VentAC (p)
Hamilton T1PCV+

Assist-control with a pressure target

Bilevel mode name
ReVelBiPAP
ZOLL Z VentBL
Hamilton T1DuoPAP

Two pressure levels — IPAP and EPAP

Adaptive mode name
ReVel—
ZOLL Z Vent—
Hamilton T1ASV

Hamilton only

Volume-guaranteed mode name
ReVel—
ZOLL Z Vent—
Hamilton T1APVcmv

Hamilton only — you set the volume and the machine finds the pressure

Airway pressure release mode name
ReVel—
ZOLL Z Vent—
Hamilton T1APRV

Hamilton only — a sustained inflation with a timed release

Dictionary

PEEP

Setting

Positive End-Expiratory Pressure

Pressure maintained in the airway at end-expiration. Recruits collapsed alveoli, improves oxygenation, and prevents atelectrauma. High PEEP can reduce venous return and cause hypotension.

FiO2

Setting

Fraction of Inspired Oxygen

The percentage of oxygen in the delivered gas (21% = room air). Titrate to the target SpO2 — use the lowest FiO2 that achieves it to avoid oxygen toxicity.

Vt

Setting

Tidal Volume

The volume of gas delivered per breath. Set on predicted (ideal) body weight, never on the weight on the stretcher — 6 mL/kg as the default for every ventilated patient, with 4–8 mL/kg the range, stepped down in ARDS while the plateau and driving pressures stay inside their limits.

Vt/IBW

Setting

Tidal Volume per Ideal Body Weight

Tidal volume normalized to ideal body weight. Hamilton expresses tidal volume this way so you set mL/kg rather than an absolute volume.

MV

Measurement

Minute Ventilation

Volume of gas moved per minute = tidal volume × respiratory rate. The primary determinant of CO2 clearance.

Ppeak

Measurement

Peak Airway Pressure

Highest pressure during inspiration. Reflects airway resistance plus lung compliance; a rising Ppeak with a normal plateau suggests a resistance problem (secretions, bronchospasm, kinked tube).

Pplat

Measurement

Plateau Pressure

Alveolar pressure at end-inspiration during a pause. Reflects lung compliance — keep ≤ 30 cmH2O to limit barotrauma.

Driving Pressure

Measurement

Pplat − PEEP

The pressure distending the lung each breath. Strongly associated with mortality in ARDS — keep ≤ 15 cmH2O.

Compliance

Mechanics

Lung Compliance

Change in volume per unit pressure (mL/cmH2O). Low compliance = stiff lungs (ARDS, pulmonary edema, pneumothorax).

Resistance

Mechanics

Airway Resistance

Opposition to airflow (cmH2O/L/s). Rises with bronchospasm, secretions, and small-bore tubes.

I:E

Setting

Inspiratory to Expiratory Ratio

Time spent inspiring versus exhaling. Normally ~1:2; lengthened toward 1:4 in obstructive disease to allow full exhalation.

TI

Setting

Inspiratory Time

Duration of the inspiratory phase in seconds. Longer TI raises mean airway pressure; shorter TI shortens the time available to exhale.

Flow

Setting

Inspiratory Flow

Rate at which gas is delivered (L/min). Higher flow shortens inspiratory time — useful in obstructive disease.

Trigger

Setting

Sensitivity

How much effort the patient must generate to start a breath. Set too sensitive → auto-triggering; too insensitive → the patient fights the vent.

PS

Setting

Pressure Support

Extra inspiratory pressure added to spontaneous breaths to overcome tube resistance and reduce work of breathing.

ΔPinsp

Setting

Inspiratory Pressure

Pressure applied above PEEP in pressure-control ventilation. Determines the tidal volume delivered.

IPAP

Setting

Inspiratory Positive Airway Pressure

The higher pressure in bilevel (BiPAP) therapy — supports inspiration and augments tidal volume.

EPAP

Setting

Expiratory Positive Airway Pressure

The lower pressure in bilevel therapy; functionally equivalent to PEEP — it recruits and splints the upper airway.

CPAP

Mode

Continuous Positive Airway Pressure

A single constant pressure throughout the breath while the patient breathes spontaneously. Splints the airway and recruits alveoli.

BiPAP

Mode

Bilevel Positive Airway Pressure

Two pressures (IPAP/EPAP) with spontaneous triggering. First-line NIV for COPD exacerbation, pulmonary edema and obesity hypoventilation.

A/C

Mode

Assist-Control

Every breath — patient-triggered or time-triggered — receives full ventilator support. Guarantees minute ventilation at the cost of hyperventilation if the patient triggers rapidly.

SIMV

Mode

Synchronized Intermittent Mandatory Ventilation

A set number of mandatory breaths are synchronized with effort; spontaneous breaths in between receive pressure support. Largely a weaning mode.

ASV

Mode

Adaptive Support Ventilation

The ventilator continuously adjusts rate and tidal volume to a target minute ventilation using the patient's measured compliance and resistance.

APRV

Mode

Airway Pressure Release Ventilation

A high continuous pressure with brief timed releases. Recruits and maintains lung volume while allowing spontaneous breathing.

PRVC

Mode

Pressure-Regulated Volume Control

A pressure breath whose pressure is adjusted breath by breath to deliver the tidal volume you set. It reconciles the comfort of pressure control with the guaranteed volume of volume control — you set the volume, the machine works out the pressure from the patient's own compliance.

Auto-PEEP

Mechanics

Intrinsic PEEP

Trapped gas raising end-expiratory alveolar pressure above the set PEEP. Common in obstructive disease; detect with an expiratory hold.

Dead Space

Mechanics

Physiologic Dead Space

Ventilated lung that does not participate in gas exchange (~2 mL/kg, roughly 150 mL in adults). It dilutes the effective alveolar ventilation.

EtCO2

Measurement

End-Tidal Carbon Dioxide

CO2 measured at end-expiration; approximates PaCO2. Guides ventilation targets and confirms endotracheal tube placement.

SpO2

Measurement

Peripheral Oxygen Saturation

Oxygen saturation estimated by pulse oximetry. Affected by poor perfusion, motion and dyshemoglobinemias.

Lung-Protective Ventilation

Strategy

Low Tidal Volume Strategy

6 mL/kg predicted body weight (4–8 mL/kg range), plateau ≤ 30, driving pressure ≤ 15, and enough PEEP to minimize volutrauma, barotrauma and atelectrauma. This is the default for every ventilated patient, not only for ARDS — prehospital and emergency ventilation is characteristically injurious, and ventilator-induced lung injury develops early.

Permissive Hypercapnia

Strategy

Tolerated Hypercapnia

Accepting an elevated CO2 (pH ≥ 7.20) in exchange for protecting the lung with low tidal volumes. It is expected with a low tidal volume, but it has two exceptions: a raised ICP, where CO2 dilates the cerebral vessels and raises the pressure inside the skull, and a sickle cell crisis, where acidosis promotes sickling. Neither patient gets a permissive CO2 target.

DOPES

Troubleshooting

Deterioration Mnemonic

Displacement of the tube, Obstruction, Pneumothorax, Equipment failure, Stacking of breaths — the checklist for sudden deterioration on a ventilator.

RSI

Airway

Rapid Sequence Intubation

Simultaneous sedation and neuromuscular blockade to secure the airway quickly and reduce aspiration risk.

BVM

Airway

Bag-Valve-Mask

Manual ventilation with a self-inflating bag. A bridge to a definitive airway — it cannot provide controlled PEEP or precise volumes.

Shunt

Mechanics

Intrapulmonary Shunt

Blood that crosses the lung without picking up oxygen. It is why a flooded or consolidated lung stays hypoxemic however high the FiO2 goes — the answer is opening the alveoli up, not adding oxygen.

Mean Airway Pressure

Mechanics

Average Pressure Over the Breath

The average pressure in the airway across a whole breath. It recruits alveoli and improves oxygenation, but it is also the pressure that opposes venous return — raise it carefully in a patient with low preload.

P/F Ratio

Measurement

PaO2 ÷ FiO2

Arterial oxygen divided by the fraction of inspired oxygen — the standard measure of how severe the oxygenation failure is. Under 300 suggests lung injury; under 100 is severe.

Predicted Body Weight

Setting

IBW — Ideal Body Weight

The weight calculated from height and sex, and the basis for every tidal volume. It is not the patient's actual weight — a 150 kg patient and a 60 kg patient of the same height get the same tidal volume.

Volutrauma

Mechanics

Over-distension Injury

Lung injury from too large a tidal volume stretching the open alveoli. This is why tidal volume is set on predicted body weight rather than the patient's actual size.

Barotrauma

Mechanics

Pressure Injury

Lung injury from excessive airway pressure — air leaks into the pleural space and a pneumothorax follows. Plateau pressure is the value to watch.

Atelectrauma

Mechanics

Repeated Opening Injury

Injury from alveoli opening and collapsing with every breath. Enough PEEP holds them open and prevents it.

Recruitment Maneuvers

Strategy

Sustained Inflation

A deliberate sustained inflation — a breath held at a high pressure for 30–40 seconds — intended to open collapsed alveoli. Prolonged recruitment maneuvers are no longer recommended: they do not improve survival and can cause harm, including barotrauma and hypotension. Open the lung with PEEP and let the FiO2 come down instead.

Prone Positioning

Strategy

Ventral Position

Turning the patient face down. In moderate–severe ARDS, prone positioning for more than 12 hours a day improves survival — it redistributes the lung's blood flow toward the dorsal units that stay open and lets the ventral units drain. It is an escalation for a P/F ratio that will not hold on PEEP and FiO2 alone, and it is not a transport intervention.

For training & education only. Not for patient care. This project is a work in progress.ContactPrivacy PolicyTerms of Use